Record the readings of one or two DMMs and keep them as a CSV file.
- Connect. Press Connect on a card and choose your DMM in the dialog of the browser. Use the second card for a second DMM.
- Choose. Give the DMM a name and choose what it measures. The range is automatic.
- Start. Choose the interval and press Start. The readings are drawn as they come. Press Stop to end.
- Save. Press Save CSV. The file keeps the time of every reading and can be opened in a spreadsheet.
Advanced settings
Each card has settings that are closed at first: the integration time (NPLC) and the range. Press Apply, and the card shows which of them the DMM accepted and what the DMM has now. A setting that is refused goes back to its default.
- A DMM that is slow (for example 0.4 seconds for a reading) can be made faster with a smaller NPLC. Fewer cycles also mean more noise.
- A fixed range saves the time the DMM needs to change the range. A reading over the range is then shown as OVLD.
- Not every DMM has an NPLC setting, and each DMM accepts only some values. The DMM tells you when it does not.
- Setting up a measurement again puts the DMM back to its defaults, so the settings are cleared when you choose another measurement.
The chart
The chart shows the last minute. Choose how much it shows with Show. While logging, it follows the newest readings. Move the slider to look at earlier readings, and press Live to go back to the newest. The slider also moves through a CSV file that you open.
Looking at a file
Open CSV shows a file of the logger again, with the smallest, the largest and the mean of each DMM. No DMM is needed.
Good to know
- The readings are kept only in this page. Save them before you close the page or reload it.
- The DMM sets how fast it measures (the integration time). If it needs more time than the interval, the page skips readings and tells you. Use a longer interval.
- The time of a reading is the clock of this computer. The two DMMs are asked together, so their readings are a few milliseconds apart.
- A reading over the range is left empty in the file.
For a long recording
- Keep the page in front, and do not close the tab. A page that is hidden or asleep may stop logging.
- Set the screen and the computer not to sleep. This page does not do that for you.
- On a phone: connect it to a charger, turn the screen timeout off, and lower the brightness.
Tested devices
- Keysight U2741A
- Keysight 34405A
Other Keysight (Agilent) DMMs may work too. Only instruments of Keysight (Agilent) are offered in the dialog of the browser. A modular DMM of the U27xx series must be switched to USBTMC mode first.